Electrotactile Literature Synthesis

What the electrotactile literature reports, and what it leaves out

A curated corpus of electrotactile, electrocutaneous and stimulator-hardware papers, each classified by what it contributes and by what stimulation parameters a paper of that kind owes its readers. Doses are recomputed on a common basis so configurations can be compared across studies.

Dose landscape

Charge per phase against charge density per phase, one point per delivered configuration. Electrode area is the free variable: points move along the diagonal as area changes.
Shaded bands: I <10, II 10–100, III >100 µC cm⁻². The dashed line is the Shannon (1992) k = 1.5 curve, derived for implanted electrodes and drawn as an external reference, not a skin limit. Values at the upper end of each paper's delivered range. Where a paper reports only ranges (Duente 2023), it is plotted at both its median-charge and its upper-end configuration, derived from the paper's own amplitude regression; the tooltip names each configuration.

Reporting completeness

Share of records reporting each parameter, scored only against what that kind of paper owes (its obligation tier).

Where this literature is published

Top three journals and conferences by number of papers in the corpus. Ties share a rank, so a list can run past three.

Corpus by contribution and obligation

Papers by primary contribution class (rows) and dose-reporting tier (columns).